Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. 2015, 159(4):527-534 | DOI: 10.5507/bp.2015.007
Glucose concentrations in blood and tissue - a pilot study on variable time lag
- a Department of Physiology, Faculty of Medicine and Dentistry, Palacky University Olomouc, Czech Republic
- b Department of Internal Medicine II - Gastroenterology and Hepatology, Faculty of Medicine and Dentistry, Palacky University Olomouc and University Hospital Olomouc
- c Institute of Neurology and Geriatrics, Moravsky Beroun
- d BVT Technologies, a.s., Strazek, Czech Republic
- e Probe Scientific Ltd, Bedford, United Kingdom
- f Heat Transfer and Fluid Flow Laboratory, Faculty of Mechanical Engineering, University of Technology, Brno
- g Department of Medical Chemistry and Biochemistry, Faculty of Medicine and Dentistry, Palacky University Olomouc
Aim: The aim of this pilot study was to acquire insight into the parameters of glycaemic control, especially, (1) the time delay (lag phase) between plasma and tissue glucose concentrations in relation to rise and fall in glucose levels and (2) the rate of glucose increase and decrease.
Methods: Four healthy people (HP), 4 people with type 1diabetes (DM1) and 4 with type 2 diabetes (DM2) underwent concurrent glucose measurements by means of (1) the continuous glucose monitoring system (CGMS-Medtronic), Medtronic-Minimed, CA, USA, calibrated by the glucometer Calla, Wellion, Austria, and, (2) the Beckman II analyser to measure glucose concentrations in venous plasma. Samples were taken on 4 consecutive days in the fasting state and 4 times after consumption of 50 g glucose. Carelink Personal, MS Excel, Maple and Mat lab were applied to plot the evolution of glucose concentration and analyse the results. The time difference between increase and decrease was calculated for HP, DM 1 and DM 2.
Results: In DM1and DM2, glucose tolerance testing (GTT) resulted in slower transport of glucose into subcutaneous tissue than in HP where the lag phase lasted up to 12 min. The maximum increase/decrease rates in DM1 and DM2 vs HP were 0.25 vs < 0.1 mmol/L/min.
Conclusion: CGMS is shown to provide reliable plasma glucose concentrations provided the system is calibrated during a steady state. The analysis of glucose change rates improves understanding of metabolic processes better than standard GTT.
Keywords: diabetes mellitus, continuous glucose monitoring, lag phase, glucose transport, calibration
Received: October 1, 2014; Accepted: January 28, 2015; Prepublished online: March 1, 2015; Published: December 3, 2015 Show citation
| ACS | AIP | APA | ASA | Harvard | Chicago | Chicago Notes | IEEE | ISO690 | MLA | NLM | Turabian | Vancouver |
References
- Takahashi S, Uchino H, Shimizu T, Kanazawa A, Tamura Y, Sakai K, Watada H, Hirose T, Kawamori R, Tanaka Y. Comparison of glycated albumin (GA) and gycated hemoglobin (HbA1c) in type 2 diabetic patients: usefulness of GA for evaluation of short-term changes in glycemic control. Endocr J 2007;54(1):139-44.
Go to original source...
Go to PubMed... - Dungan KM. 1,5-anhydroglucitol (GlycoMark™) as a marker of short-term glycemic control and glycemic excursions. Expert Rev Mol Diagn 2008;8(1):9-19.
Go to original source...
Go to PubMed... - Rodbard D. Interpretation of continuous glucose monitoring data: glycemic variability and quality of glycemic control. Diabetes Technol Ther 2009;11(Suppl 1):S55-S65.
Go to original source...
Go to PubMed... - Kohnert K-D, Heinke P, Vogt L, Meyramov GG, Salzsieder E. Optimizing diabetes management: comprehensive analysis of glucose monitoring data and use of better metrics for glycemic control. Bulletin of the Karaganda University 2013;4(72):6-15.
Go to original source... - Peterson K, Chlup R, Zapletalová J, Kohnert K, Kudlová P, Bartek J, Nakládalová M, Doubravová B, Sečkař P. Influence of oral antidiabetic drugs on hyperglycemic response to foods in persons with type 2 diabetes mellitus as assessed by continuous glucose monitoring system:a pilot study. J Diab Sci Tech 2010;4(4):983-92.
Go to original source...
Go to PubMed... - Chlup R, Doubravová B, Bartek J, Zapletalová J, Krystyník O, Procházka V. Effective assessment of diabetes control using personal glucometers (Contourlink, Bayer, Germany; Calla, Wellion, Austria; Linus, Agamatrix, USA). Disease Markers 2013;35(6):895-905.
Go to original source...
Go to PubMed... - Mlčák P, Fialová J, Trnková K, Chlup R. A continuous glucose monitoring system (CGMS) - a promising aproach for improving metabolic control in persons with type 1 diabetes mellitus treated by insulin pumps. Biomed Papers 2004;148(1):33-8.
Go to original source... - Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group, Bode B, Beck RW, Xing D, Gilliam L, Hirsch I, Kollman C, Laffel L, Ruedy KJ, Tamborlane WV, Weinzimer S, Wolpert H. Sustained benefit of continuous glucose monitoring on HbA1c, glucose profiles, and hypoglycemia in adults with type 1 diabetes. Diabetes Care 2009;32(11):2047-9.
Go to original source...
Go to PubMed... - Cohen O, Körner A, Chlup R, Zoupas C, Rahozin A, Wudi K, Bartášková D, Pappas A, Niederland T, Taybani Z, Barák L, Vazeou AS. Improved glycemic control through continuous glucose sensor - augmented insulin pump therapy: Prospective results from a community and academic practice patient registry. J Diab Sci Techn 2009;3(4):804-11.
Go to original source...
Go to PubMed... - Peterson K, Zapletalová J, Kudlová P, Matušková V, Bartek J, Novotný D, Chlup R. Benefits of three-month continuous glucose monitoring for persons with diabetes using insulin pumps and sensors. Biomed Papers 2009;153(1):47-52.
Go to original source... - Battelino T, Conget I, Olsen B, Schütz-Fuhrmann I, Hommel E, Hoogma R, Schierloh U, Sulli N, Bolinder J; SWITCH Study Group. The use and efficacy of continuous glucose monitoring in type 1 diabetes treated with insulin pump therapy: a randomised controlled trial. Diabetologia 2012;55:3155-62.
Go to original source...
Go to PubMed... - Vigersky RA, Fonda SJ, Chellappa M, Walker MS, Ehrhardt MN. Short- and long-term effects of real-time continuous glucose monitoring in patients with type 2 diabetes. Diabetes Care 2012;35(1):32-8.
Go to original source...
Go to PubMed... - Poljaková I, Elšíková E, Chlup R, Kalabus S, Hasala O, Zapletalová J. Glucose sensing module - is it time to integrate it into real-time perioperative monitoring? An observational pilot study with subcutaneous sensors. Biomed Pap Med Fac Univ Palacky Olomouc Czech Rep 2013;157(4):346-57.
Go to original source...
Go to PubMed... - Chlup R, Jelenová D, Kudlová P, Chlupová K, Bartek J, Zapletalová J, Langová K, Chlupová L. Continuous glucose monitoring - novel approach to the determination of the glycaemic index of foods (DEGIF 1); Determination of the glycaemic index of foods by means of the CGMS. Exp Clin Endocrinol Diab 2006;144(12):68-74.
Go to original source...
Go to PubMed... - Chlup R, Peterson K, Zapletalová J, Kudlová P, Sečkař P. Extended prandial glycemic profiles of foods as assessed using continuous glucose monitoring enhance the power of the 120-minute glycemic index. J Diab Sci Tech 2010;4(3):615-24.
Go to original source...
Go to PubMed... - Rebrin K, Steil GM. Can Interstitial Glucose Assessment Replace Blood Glucose Measurements? Diabetes Technology & Therapeutics. October 2000;2(3):461-72. doi:10.1089/15209150050194332.
Go to original source... - Thennadil SN, Rennert JL, Wenzel BJ, Hazen KH, Ruchti TL, Block MB. Comparison of glucose concentration in interstitial fluid, and capillary and venous blood during rapid changes in blood glucose levels. Diabetes Technology & Therapeutics. 2001;3(3):357-65. doi:10.1089/15209150152607132.
Go to original source... - Boyne MS, Silver DM, Kaplan J, Saudek CD. Timing of changes in interstitial and venous blood glucose measured with a continuous subcutaneous glucose sensor. Diabetes 2003;52(11):2790-4. doi: 10.2337/diabetes.52.11.2790.
Go to original source...
Go to PubMed... - Hovorka R, Nodale M. Response to Mitre et al.: Analysis of Continuous Glucose Monitoring Data to Assess Outpatient Closed-Loop Studies: Considerations for Different Sensors. Diabetes Technology & Therapeutics 2014;16(5):328-9. doi:10.1089/dia.2013.0336.38.
Go to original source... - Leelarathna L, Dellweg S, Mader JK, Allen JM, Benesch C, Doll W, Ellmerer M, Hartnell S, Heinemann L, Kojzar H, Michalewski L, Nodale M, Thabit H, Wilinska ME, Pieber TR, Arnolds S, Evans ML, Hovorka R; AP@home Consortium. Day and night home closed-loop insulin delivery in adults with type 1 diabetes: three-center randomized crossover study. Diabetes Care 2014; 37(7):1931-7.
Go to original source...
Go to PubMed... - Bergenstal RM, Ahmann AJ, Bailey T, Beck RW, Bissen J, Buckingham B, Deeb L, Dolin RH, Garg SK, Goland R, Hirsch IB, Klonoff DC, Kruger DF, Matfin G, Mazze RS, Olson BA, Parkin C, Peters A, Powers MA, Rodriguez H, Southerland P, Strock ES, Tamborlane W, Wesley DM. Recommendations for standardizing glucose reporting and analysis to optimize clinical decision making in diabetes: the Ambulatory Glucose Profile (AGP). Diabetes Technol Ther 2013;15(3):198-211. doi: 10.1089/dia.2013.0051.
Go to original source...
Go to PubMed... - Chlup R, Krystynik O, Nadvornikova M, Zalesakova M, Durajkova E, Poljakova I, Kudlova P, Bartek J, Zapletalova J, Prochazka V. Benefits of advanced technologies for optimum therapy of type 1 and type 2 diabetes in practice. (Moderní technologie a způsob jejich využívání pro optimalizaci léčby diabetu 1. i 2. typu v praxi). Klin Farmakol Farmacie 2014;28(2):72-9.
- Rodbard D. Clinical interpretation of indices of quality of glycemic control and glycemic variability. Postgrad Med 2011;123(4):107-18.
Go to original source...
Go to PubMed... - Salzsieder E, Vogt L, Heinke P, Augstein P. Q-score: a new tool for objective assessment and verification of the quality of continuously measured glucose profiles. Diabetologie und Stoffwechsel 2011;6:223.
Go to original source... - Kohnert KD, Vogt L, Salzsieder E. Advances in understanding glucose variability and the role of continuous glucose monitoring. European Endocrinol 2010;6(1):53-56.
Go to original source... - Fritzsche G, Kohnert KD, Heinke P, Vogt L, Salzsieder E. The use of a computer program to calculate the mean amplitude of glycemic excursions. Diabetes Technol Ther 2011;13(3):319-25.
Go to original source...
Go to PubMed... - Kohnert KD, Augstein P, Heinke P, Zander E, Peterson K, Freyse EJ, Salzsieder E. Chronic hyperglycemia but not glucose variability determines HbA1c levels in well-controlled patients with type 2 diabetes. Diabetes Res Clin Pract 2007;77(3):420-6.
Go to original source...
Go to PubMed... - Kohnert KD, Augstein P, Zander E, Heinke P, Peterson K, Freyse EJ, Hovorka R, Salzsieder E. Glycemic variability correlates strongly with postprandial beta-cell dysfunction in a segment of type 2 diabetic patients using oral hypoglycemic agents. Diabetes Care 2009;32(6):1058-62.
Go to original source...
Go to PubMed... - Rawlings RA, Shi H, Yuan LH, Brehm W, Pop-Busui R, Nelson PW. Translating glucose variability metrics into the clinic via continuous glucose monitoring: a graphical user interface for diabetes evaluation (CGM-GUIDE©)// Diabetes Technol Ther 2011;13(12):1241-8.
Go to original source... - Rutscher A, Salzsieder E, Thierbach U, Fischer U, Albrecht G. KADIS - a computer-aided decision support system for improving the management of type-I diabetes. Exp Clin Endocrinol 1990;95(1):137-47.
Go to original source...
Go to PubMed... - Augstein P, Vogt L, Kohnert KD, Freyse EJ, Heinke P, Salzsieder E. Outpatient assessment of Karlsburg Diabetes Management System-based decision support. Diabetes Care 2007;30(7):1704-8.
Go to original source...
Go to PubMed... - Salzsieder E, Vogt L, Kohnert KD, Heinke P, Augstein P. Model-based decision support in diabetes care. Comput Methods Prog Biomed 2011;102(2):206-18.
Go to original source...
Go to PubMed... - Rodbard D, Vigersky RA. Design of a decision support system to help clinicians manage glycemia in patients with type 2 diabetes mellitus. J Diabetes Sci Technol 2011;5(2):402-11.
Go to original source...
Go to PubMed... - Mastrototaro J, Welsh JB, Lee S. Practical considerations in the use of real-time continuous glucose monitoring alerts. J Diabetes Sci Technol 2010;4(3):733-9.
Go to original source...
Go to PubMed... - Reznik Y, Cohen O, Aronson R, Conget I, Runzis S, Castaneda J, Lee SW. Insulin pump treatment compared with multiple daily injections for treatment of type 2 diabetes (OpT2mise): a randomised open-label controlled trial. The Lancet. Jul 2014 E pub ahead of print doi: 10.1016/S0140-6736(14)61037-0.
Go to original source...