Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. 2007, 151(2):263-266 | DOI: 10.5507/bp.2007.044
OCCURENCE OF ADVERSE EVENTS DUE TO CONTINUOUS GLUCOSE MONITORING
- a Department of Physiology and 2nd Department of Medicine, Faculty of Medicine and Dentistry, Palacky University, Olomouc, Czech Republic
- b Department of Physiology and 1st Department of Medicine, Faculty of Medicine, P. J. Safarik University, Kosice, Slovak Republic
Aims: Continuous glucose monitoring (CGM) using transcutaneous sensors is becoming a sophisticated method to control and regulate glucose metabolism. The transcutaneous sensor of the CGM system (CGMS™ Medtronic Minimed, Northridge, CA, USA) is chosen to measure glucose concentration in interstitial fluid up to three days after insertion even though its function remains stable for a longer period. The question arises, which factors really limit the period of sensor insertion without unnecessary risk. The aim of this study was to assess any adverse events occurring in the course of 9 days after the sensor insertion.
Methods: In a group of 22 healthy volunteers aged 21.8±1.30 y (mean ± SE) a total of 26 sensors was inserted subcutaneously in gluteal or lumbar region for 9 days. Before insertion the site was sprayed with an antiseptic (Cutasept F, Bode Chemie, Hamburg, Germany). Local adverse reactions and disturbances in general condition were examined.
Results: In the course of 184 sensor-days, there were only minor local adverse events: hypersensitivity, itching, pain, redness, burning, subcutaneous hemorrhage. Additionally, sleep disturbances, attention deficits, problems related to the CGMS monitor, to adhesive tape and/or sensor were found. None of these resulted in sensor withdrawal. In 12 volunteers (55 %) no complications were observed. The sensor function measured according to electrical signals (ISIG) failed (always on day 1-2) in 4 cases (16 %).
Conclusions: The present FDA approved 3-day insertion period for Medtronic transcutaneous sensor does not seem to limit its use and appears to be worth a careful revision.
Keywords: Continuous glucose monitoring, Sensors, Adverse events
Received: September 12, 2007; Accepted: October 18, 2007; Published: December 1, 2007 Show citation
References
- Mastrototaro JJ. The MiniMed Continuous Glucose Monitoring System (CGMS). J Pediatr Endocr Metab. 1999;12:751758.
- Klonoff DC. A review of continuous glucose monitoring technology. Diabetes Technol Ther. 2005; 7:770775.
Go to original source...
- Mastrototaro JJ, Cooper KW, Sounararajan G, Sanders JB, Shah RV. Clinical experience with an integrated continuous glucose sensor/ insulin pump platform: a feasibility study. Adv Ther [serial on the internet]. 2006 Sep-Oct [cited 2007 Sep 6];23(5):72532.
Go to original source...
Go to PubMed...
- Garg SK, Schwartz S, Edelman SV. Improved glucose excursions using an implantable real-time continuous glucose sensor in adults with type 1 diabetes. Diabetes Care. 2004; 27:734738.
Go to original source...
Go to PubMed...
- Garg S, Zisser H, Schwartz S, Bailey T, Kaplan R, Ellis S et al. Improvement in glycemic excursions with a transcutaneous, realtime continuous glucose sensor: a randomized controlled trial. Diabetes Care. 2006; 29:4450.
Go to original source...
Go to PubMed...
- Buckingham B, Kunselman B, Wong L, Istok E, Leach J, Purvis R. Extended use of a new glucose sensor with wireless data transmission: the next generation CGMS [abstract 243]. Diabetologia. 2004; 47(Suppl 1):A93.
- Mlcak P, Fialova J, Trnkova K, Chlup R. A continous glucose monitoring system (CGMS) A promising approach for improving metabolic control in persons with type 1 diabetes mellitus treated by insulin pumps. Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. 2004; 148:3338.
Go to original source...
Go to PubMed...
- Chlup R, Jelenova D, Chlupova K, Zapletalova J, Chlupova L, Bartek J. Function and Accuracy of Glucose Sensors beyond their stated Expiry Date. Diabetes Technology & Therapeutics. 2005; 8(4):495504.
Go to original source...
- Chlup R, Jelenová D, Kudlová P, Chlupová K, Bartek J, Zapletalová J et al. Continuous glucose monitoring a novel approach to the determination of the glycaemic index of foods (DEGIF 1). Exp Clin Endocrinol Diabetes. 2006; 114:6874.
Go to original source...
Go to PubMed...
- Chlupová K, Kohnert KD, Heinke P, Augstein P, Chlup R, Salzsieder E. Resultate kontinuierlicher Glukosemessungen (CGMS) unter ambulanten Bedingungen zur Beurteilung der Stoff wechseleinstellung bei T2DM. Diabetologie und Stoff wechsel. 2006; 1(Suppl 1):94(Abstract).
Go to original source...
- Chlup R, Kudlova P, Seckar P, Zapletalova J, Bartek J, Chlupova K, Luza J. Glycaemic index of foods in healthy persons vs persons with type 1 diabetes mellitus. Diabetes 2006; 55(Suppl 1): A593(Abstract).
- Chlup R, Payne M, Zapletalova J, Komenda S, Doubravova B, Reznickova M, Chlupova L, Seckar P. Results of selfmonitoring on glucometer systems Advance and Optium in daily routine. Biomed Pap. 2005;149:127139.
Go to original source...
- Bode BW. Clinical utility of the Continuous Glucose Monitoring System. Diabetes Technol Ther. 2000;2:3541.
Go to original source...
- Rebrin K, Steil GM. Can interstitial glucose assessment replace blood glucose measurements? Diabetes Technol Ther. 2000; 2:461 471.
Go to original source...
Go to PubMed...
- Bode BW, Gross TM, Thornton KR, Mastrototaro JJ. Continuous glucose monitoring used to adjust diabetes therapy improves glycosylated hemoglobin: a pilot study. Diabetes Res Clin Pract. 1999; 46:183190.
Go to original source...
Go to PubMed...
- Broz J, Andel M. Continuous glucose monitoring can help to decrease frequency of syptomatic hypoglycaemia in patients with type 1 diabetes treated with CSII in long term follow-up. Diabetes. 2002; 45:480.
- De Block CEM, Keenoy BM, Van Gaal LF, Rogiers P. Continuous glucose monitoring in a medical intensive care unit. Diabetologia. 2005; 48:A48.
- Donicova V, Pallayova M, Donic V. Clinical utility of the continuous glucose monitoring system in intensifi ed insulin treated patients. Diabetologia. 2005; 48:A50.
- Chlup R, Seckar P, Zapletalova J, Langova K, Kudlova P, Peterson K, Bartek J, Hucikova J. DegifXL a new software program for determination of group-related and person-related glycemic indexes of foods by means of continuous glucose monitoring system CGMS. Acta Diab. Romana. 2007; 33/1:351.
- Gross TM, Mastrototaro JJ. Effi cacy and reliability of the continuous glucose monitoring system. Diabetes Technol Ther [serial on the internet]. 2000 [cited 2007 Sep 6];2 (Suppl 1):S1926.
Go to original source...
- Gross TM, Bode BW, Einhorn D, KayneDM, Reed JH, White NH, Mastrotorato JJ. Performance evaluation of the MiniMed continuous glucose monitoring system during patient home use. Diabetes Technol Ther. 2000; 2(1):4956.
Go to original source...
- Chlup R, Janů K, Venháčová J, Bartek J. Six models of a new insulin pen (MADI): description and fi rst clinical trial. Practical Diabetes International. 1995; 12(1).
Go to original source...
- Chlup R, Marąálek E, Bruns V. A prospective study of the hazards of multiple use of disposable syringes ane needles in intensifi ed insulin therapy. Diabetic Medicine. 1990; 7: 624627.
Go to original source...
Go to PubMed...
- Lippaiova N, Pallayova M, Peterson K, Luza J, Fajkosova L. Safety of novel algorithms for premeal insulin boluses in high glycaemic index meals in persons with type 1 diabetes mellitus using insulin pumps. Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. Submitted.
- Fajkusova Z, Jadviscokova T, Pallayova M, Matuskova V, Luza J, Kuzmina G. Glycaemic index of selected foodstuff s in healthy persons. Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. 2007;151(2).
Go to original source...
Go to PubMed...
- Pribylova H, Pallayova M, Hucikova J, Luza J. Evaluation of the new software programme DegifXL4 in the the determination of the glycaemic indices of foodstuff s. Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. Submitted.